Genesys 30 |
Genesys 40 |
Genesys 140 |
Genesys 50 |
Genesys 150 |
BioMate 160 |
Genesys 180 |
|
Optical Design |
Single beam |
Dual beam |
Double beam |
|||||
Spectral bandwidth |
5 nm |
2 nm |
||||||
Source |
Tungsten - halogen |
Xenon |
||||||
Detector |
Silicon photodiode |
Dual Silicon photodiode |
||||||
Spectral range |
325 – 1100 nm |
190 – 1100 nm |
||||||
Wavelength accuracy |
± 2 nm |
± 0.5 nm |
||||||
Wavelength repeatability |
< ± 1 nm |
< ± 0.2 nm |
||||||
Scan speed |
Up to 1200 nm/min |
Up to 1800 nm/min |
Up to 1600 nm/min |
|||||
Data resolution |
1 nm |
0.2; 0.5; 1; 2; 5 nm |
0.1; 0.2; 0.5; 1; 2; 5 nm |
|||||
Photometric range |
-3A - +3A |
-3A – +3.5 A |
-2A – +3.5 A |
|||||
Photometric accuracy |
± 0.002 A (0-0.3A) |
± 0.002 A at 0.5 A; ± 0.004 A at 1.0 A; ± 0.008 A at 2.0 A |
||||||
Photometric repeatability |
± 0.002 A |
± 0.001 A at 1.0 A |
||||||
Photometric noise |
≤0.001A at 0;1.0 A |
≤ 0.0002 at 0A; ≤ 0.0003 at 1A; ≤ 0.0004 at 2A (260 at 500 nm) |
||||||
Drift |
< 0.002 A/Hr |
< 0.0010 A/Hr |
< 0.0005 A/Hr |
|||||
Stray light |
<0.1%T (340 and 400 nm) |
<0.05%T (340 and 400 nm) |
< 1.0%T 198 nm; < 0.05%T 220 nm; < 0.03%T 340 nm |
|||||
Baseline flatness |
<0.003A |
±0.005A |
±0.002A |
|||||
Display |
5 inch |
7 inch touchscreen |
||||||
Interface |
USB printer/PC |
|||||||
Control |
Local, integrated software and/or computer with external software |